#!/usr/bin/env python3
|
|
# -*- coding: utf-8 -*-
|
|
|
|
#
|
|
# This application demonstrates how to communicate with an oximeter
|
|
# using HDP. This application has been tested against a
|
|
# Nonin Onyx II 9560 pulse oximeter.
|
|
#
|
|
|
|
import sys
|
|
import os
|
|
from gi.repository import GLib
|
|
import dbus
|
|
import socket
|
|
import dbus.service
|
|
from gi.repository import GObject
|
|
from dbus.mainloop.glib import DBusGMainLoop
|
|
from dbus.exceptions import DBusException
|
|
import argparse
|
|
|
|
# from hdp_utils import *
|
|
from random import randint
|
|
|
|
BUS_NAME = 'org.bluez'
|
|
PATH = '/org/bluez'
|
|
HEALTH_MANAGER_INTERFACE = 'org.bluez.HealthManager1'
|
|
HEALTH_DEVICE_INTERFACE = 'org.bluez.HealthDevice1'
|
|
HEALTH_CHANNEL_INTERFACE = 'org.bluez.HealthChannel1'
|
|
|
|
class MessageType:
|
|
(Association, Configuration, Release_Request,
|
|
Release_Confirmation, Data, Unknown) = range(0, 6)
|
|
|
|
class HdpMessage:
|
|
def getAssociationResponse(self, invokeId):
|
|
return bytes((
|
|
0xe3, 0x00, #APDU CHOICE Type(AareApdu)
|
|
0x00, 0x2c, #CHOICE.length = 44
|
|
0x00, 0x00, #result=accept (known config)
|
|
0x50, 0x79, #data-proto-id = 20601
|
|
0x00, 0x26, #data-proto-info length = 38
|
|
0x80, 0x00, 0x00, 0x00, #protocolVersion
|
|
0x80, 0x00, #encoding rules = MDER
|
|
0x80, 0x00, 0x00, 0x00, #nomenclatureVersion
|
|
0x00, 0x00, 0x00, 0x00, #functionalUnits, normal Association
|
|
0x80, 0x00, 0x00, 0x00, #systemType = sys-type-manager
|
|
0x00, 0x08, #system-id length = 8 and value (manufacturer- and device- specific)
|
|
0x88, 0x77, 0x66, 0x55, 0x44, 0x33, 0x22, 0x11,
|
|
0x00, 0x00, #Manager's response to config-id is always 0
|
|
0x00, 0x00, #Manager's response to data-req-mode-flags is always 0
|
|
0x00, 0x00, #data-req-init-agent-count and data-req-init-manager-count are always 0
|
|
0x00, 0x00, 0x00, 0x00, #optionList.count = 0 | optionList.length = 0
|
|
))
|
|
|
|
def getConfigurationResp(self, invokeId):
|
|
return bytes((
|
|
0xe7, 0x00, #APDU CHOICE Type(PrstApdu)
|
|
0x00, 0x16, #CHOICE.length = 22
|
|
0x00, 0x14, #OCTET STRING.length = 20
|
|
invokeId[0], invokeId[1], #invoke-id (mirrored from invocation)
|
|
0x02, 0x01, #CHOICE(Remote Operation Response | Confirmed Event Report)
|
|
0x00, 0x0e, #CHOICE.length = 14
|
|
0x00, 0x00, #obj-handle = 0 (MDS object)
|
|
0x00, 0x00, 0x00, 0x00, #currentTime = 0
|
|
0x0d, 0x1c, #event-type = MDC_NOTI_CONFIG
|
|
0x00, 0x04, #event-reply-info.length = 4
|
|
0x40, 0x00, # ConfigReportRsp.config-report-id=0x4000
|
|
0x00, 0x00 # ConfigReportRsp.config-result = accepted-config
|
|
))
|
|
|
|
def getReleaseRequest(self, invokeId):
|
|
return bytes((0xe4, 0x00, 0x00, 0x02, 0x00, 0x00))
|
|
|
|
def getReleaseResponse(self, invokeId):
|
|
return bytes((0xe5, 0x00, 0x00, 0x02, 0x00, 0x00))
|
|
|
|
def getDataResponse(self, invokeId):
|
|
return bytes((
|
|
0xe7, 0x00, #APDU CHOICE Type(PrstApdu)
|
|
0x00, 0x12, #CHOICE.length = 18
|
|
0x00, 0x10, #OCTET STRING.length = 16
|
|
invokeId[0], invokeId[1], #invoke-id (mirrored from invocation)
|
|
0x02, 0x01, #CHOICE(Remote Operation Response | Confirmed Event Report)
|
|
0x00, 0x0a, #CHOICE.length = 10
|
|
0x00, 0x00, #obj-handle = 0 (MDS object)
|
|
0x00, 0x00, 0x00, 0x00, #currentTime = 0
|
|
0x0d, 0x1d, #event-type = MDC_NOTI_SCAN_REPORT_FIXED
|
|
0x00, 0x00, #event-reply-info.length = 0
|
|
))
|
|
|
|
def parse(self, string_msg):
|
|
#
|
|
# Parse a receive message. This is just example code and
|
|
# should not be used in any real application. HDP messages
|
|
# are encoded using IEEE 11073 encoding rules. Since these
|
|
# rules use ASN.1 encoding, it is possible that fields could
|
|
# move in messages. This routine does not handle that
|
|
# possibility. Instead it assumes the fields are in fixed
|
|
# locations, which they appear to be in the units we tested
|
|
# against. However, in a shipping medical device, you would
|
|
# want to handle things like this.
|
|
#
|
|
msg_type = MessageType.Unknown
|
|
invokeId = (0, 0)
|
|
sp02 = 0
|
|
pulse = 0
|
|
if debugOn:
|
|
print("IEEE opcode received: %x, length = %d" % (int(string_msg[0]), len(string_msg)))
|
|
for i in range(len(string_msg)):
|
|
if ((i & 15) == 0):
|
|
print
|
|
print('%2.2X' % int(string_msg[i]),)
|
|
print
|
|
|
|
if int(string_msg[0]) == 0xe2:
|
|
msg_type = MessageType.Association
|
|
elif int(string_msg[0]) == 0xe7:
|
|
invokeId = int(string_msg[6]), int(string_msg[7])
|
|
if int(string_msg[18]) == 0x0d and int(string_msg[19]) == 0x1c:
|
|
msg_type = MessageType.Configuration
|
|
else:
|
|
msg_type = MessageType.Data
|
|
sp02 = int(string_msg[35])
|
|
pulse = int(string_msg[49])
|
|
elif int(string_msg[0]) == 0xe4:
|
|
msg_type = MessageType.Release_Request
|
|
elif int(string_msg[0]) == 0xe5:
|
|
msg_type = MessageType.Release_Confirmation
|
|
else:
|
|
msg_type = MessageType.Unknown
|
|
|
|
return (msg_type, invokeId, sp02, pulse)
|
|
|
|
|
|
|
|
def receive_data(sk, evt):
|
|
#
|
|
# This method is called when we receive an event on our socket.
|
|
# It's probably either a message from the oximeter or a disconnect
|
|
# indicate. Deal with it.
|
|
#
|
|
data = None
|
|
disconnecting = False
|
|
hdp = HdpMessage()
|
|
if evt & GLib.IO_IN:
|
|
try:
|
|
data = sk.recv(1024)
|
|
except IOError:
|
|
data = ""
|
|
if data:
|
|
result = hdp.parse(data)
|
|
msgType = result[0]
|
|
invokeId = result[1]
|
|
if msgType == MessageType.Association:
|
|
if debugOn:
|
|
print("Oximeter has associated")
|
|
sk.send(hdp.getAssociationResponse(invokeId))
|
|
elif msgType == MessageType.Configuration:
|
|
if debugOn:
|
|
print("Received configuration data")
|
|
sk.send(hdp.getConfigurationResponse(invokeId))
|
|
elif msgType == MessageType.Release_Request:
|
|
if debugOn:
|
|
print("Received release request")
|
|
sk.send(hdp.getReleaseResponse(invokeId))
|
|
disconnecting = True
|
|
elif msgType == MessageType.Release_Confirmation:
|
|
if debugOn:
|
|
print("Received release confirmation")
|
|
disconnecting = True
|
|
elif msgType == MessageType.Data:
|
|
sk.send(hdp.getDataResponse(invokeId))
|
|
sp02 = result[2]
|
|
pulse = result[3]
|
|
if debugOn:
|
|
print("Received data from oximeter")
|
|
print("SpO2 Level: %d, Beats/second: %d" % \
|
|
(result[2], result[3]))
|
|
if debugOn:
|
|
print("Sending disconnect")
|
|
sk.send(hdp.getReleaseRequest(invokeId))
|
|
else:
|
|
print("Received unknown message, disconnecting")
|
|
sk.send(hdp.getReleaseRequest(invokeId))
|
|
disconnecting = True
|
|
|
|
if disconnecting or evt != GLib.IO_IN or not data:
|
|
try:
|
|
sk.shutdown(2)
|
|
except IOError:
|
|
pass
|
|
sk.close()
|
|
print("Disconnected from oximeter")
|
|
return False
|
|
else:
|
|
return True
|
|
|
|
|
|
class SignalHandler(object):
|
|
def __init__(self):
|
|
bus.add_signal_receiver(self.ChannelConnected,
|
|
signal_name="ChannelConnected",
|
|
bus_name=BUS_NAME,
|
|
path_keyword="device",
|
|
interface_keyword="interface",
|
|
dbus_interface=HEALTH_DEVICE_INTERFACE)
|
|
|
|
bus.add_signal_receiver(self.ChannelDeleted,
|
|
signal_name="ChannelDeleted",
|
|
bus_name=BUS_NAME,
|
|
path_keyword="device",
|
|
interface_keyword="interface",
|
|
dbus_interface=HEALTH_DEVICE_INTERFACE)
|
|
|
|
def ChannelConnected(self, channel, interface, device):
|
|
print("%s has connected" % device)
|
|
if debugOn:
|
|
print("Channel: %s" % channel)
|
|
#
|
|
# The oximeter has connected to us. Let's get
|
|
# a socket for the connection.
|
|
#
|
|
try:
|
|
channel = bus.get_object(BUS_NAME, channel)
|
|
channel = dbus.Interface(channel, HEALTH_CHANNEL_INTERFACE)
|
|
fd = channel.Acquire()
|
|
fd = fd.take()
|
|
sk = socket.fromfd(fd, socket.AF_UNIX, socket.SOCK_STREAM)
|
|
os.close(fd)
|
|
|
|
# Now set up our receiver function to be called
|
|
# when interesting events are detected on that
|
|
# socket
|
|
watch_bitmap = GLib.IO_IN | GLib.IO_ERR| GLib.IO_HUP | GLib.IO_NVAL
|
|
GLib.io_add_watch(sk, watch_bitmap, receive_data)
|
|
except DBusException:
|
|
print("Error communicating with Oximeter.")
|
|
print("Please make sure the Oximeter has fresh batteries.")
|
|
|
|
def ChannelDeleted(self, channel, interface, device):
|
|
print("Device %s channel %s deleted" % (device, channel))
|
|
|
|
parser = argparse.ArgumentParser()
|
|
parser.add_argument("-d", "--debug", help="supply debug output",
|
|
action="store_true")
|
|
args = parser.parse_args()
|
|
debugOn = args.debug
|
|
|
|
DBusGMainLoop(set_as_default=True)
|
|
loop = GObject.MainLoop()
|
|
bus = dbus.SystemBus()
|
|
|
|
signal_handler = SignalHandler()
|
|
|
|
#
|
|
# This dictionary holds the parameters that will be passed to the
|
|
# Bluez CreateApplication method. Note that we explicitly set the
|
|
# signature to indicate each entry is a string/variant pair. If
|
|
# you don't do this then you will get a cryptic error when you call
|
|
# the CreateApplication method.
|
|
#
|
|
config = dbus.Dictionary({"Role": "Sink", "DataType": dbus.types.UInt16(0x1004),
|
|
"Description": "Oximeter sink"}, signature='sv')
|
|
|
|
manager = dbus.Interface(bus.get_object(BUS_NAME, PATH),
|
|
HEALTH_MANAGER_INTERFACE)
|
|
app = manager.CreateApplication(config)
|
|
print("HDP application created, waiting for connection from")
|
|
print("a pulse oximeter. Press control-c to terminate.")
|
|
|
|
try:
|
|
loop = GLib.MainLoop()
|
|
loop.run()
|
|
except KeyboardInterrupt:
|
|
pass
|
|
finally:
|
|
manager.DestroyApplication(app)
|
|
print
|
|
print("Application stopped")
|
|
print
|